Radiation Safety Plan Supplement
Storage means a condition in which a device or source is not being used for an extended period of time, and has been made inoperable by a person registered under Sec. B.6 of this Part.
Accessible surface means the external surface of the enclosure or housing provided by the manufacturer.
Added filtration means any filtration which is in addition to the inherent filtration.
Aluminum equivalent means the thickness of type 1100 aluminum alloy 1/ affording the same attenuation, under specified conditions, as the material in question.
Assembler means any person engaged in the business of assembling, replacing, or installing one or more components into an x-ray system or subsystem. The term includes the owner of an x-ray system or his or her employee or agent who assembles components into an x-ray system that is subsequently used to provide professional or commercial services.
Attenuation block means a block or stack, having dimensions 20 centimeters by 20 centimeters by 3.8 centimeters, of type 1100 aluminum alloy 1/ or other materials having equivalent attenuation.
Automatic exposure control means a device which automatically controls one or more technique factors in order to obtain at a preselected location(s) a required quantity of radiation (See also Phototimer).
Barrier (See Protective barrier).
Beam axis means a line from the source through the centers of the x-ray fields.
1/ The nominal chemical composition of type 1100 aluminum alloy is 99.00 percent minimum aluminum, 0.12 percent copper.
Beam-limiting device means a device which provides a means to restrict the dimensions of the x-ray field.
Beam monitoring system means a system designed to detect and measure the radiation present in the useful beam.
Cephalometric device means a device intended for the radiographic visualization and measurement of the dimensions of the human head.
Certified components means components of x-ray systems which are subject to regulations promulgated under Public Law 90-602, the Radiation Control for Health and Safety Act of 1968.
Certified system means any x-ray system which has one or more certified component(s).
Changeable filters means any filter, exclusive of inherent filtration, which can be removed from the useful beam through any electronic, mechanical, or physical process.
Coefficient of variation or C means the ratio of the standard deviation to the mean value of a population of observations. It is estimated using the following equation:
Rough Equation
where
s = Estimated standard deviation of the population.
X = Mean value of observations in sample.
Xi = ith observation in sample.
n = Number of observations in sample.
Computed tomography means the production of a tomogram by the acquisition and computer processing of x-ray transmission data.
Contact therapy system means an x-ray system used for therapy with the x-ray tube port placed in contact with or within 5 centimeters of the surface being treated.
Control panel means that part of the x-ray control upon which are mounted the switches, knobs, pushbuttons, and other hardware necessary for manually setting the technique factors.
Cooling curve means the graphical relationship between heat units stored and cooling time.
CT (See Computed tomography).
Dead-man switch means a switch so constructed that a circuit closing contact can be maintained only by continuous pressure on the switch by the operator. Detector (See Radiation detector).
Diagnostic source assembly means the tube housing assembly with a beam-limiting device attached.
Diagnostic x-ray system means an x-ray system designed for irradiation of any part of the human body for the purpose of diagnosis or visualization.
Direct scattered radiation means that scattered radiation which has been deviated in direction only by materials irradiated by the useful beam (See Scattered radiation).
Entrance exposure rate means the exposure per unit time at the point where the center of the useful beam enters the patient.
Equipment (See X-ray equipment).
Field emission equipment means equipment which uses an x-ray tube in which electron emission from the cathode is due solely to the action of an electric field.
Filter means material placed in the useful beam to absorb preferentially selected radiations.
Fluoroscopic imaging assembly means a subsystem in which x-ray photons produce a fluoroscopic image. It includes the image receptor(s) such as the image intensifier and spot-film device, electrical interlocks, if any, and structural material providing linkage between the image receptor and diagnostic source assembly.
Focal spot means the area projected on the anode of the x-ray tube by the electrons accelerated from the cathode and from which the useful beam originates.
General purpose radiographic x-ray system means any radiographic x-ray system which, by design, is not limited to radiographic examination of specific anatomical regions.
Gonad shield means a protective barrier for the testes or ovaries.
Half-value layer means the thickness of specified material which attenuates the beam of radiation to an extent such that the exposure rate is reduced to one-half of its original value. In this definition, the contribution of all scattered radiation, other than any which might be present initially in the beam concerned, is deemed to be excluded.
Healing arts screening means the testing of human beings using x-ray machines for the detection or evaluation of health indications when such tests are not specifically and individually ordered by a licensed practitioner of the healing arts legally authorized to prescribe such x-ray tests for the purpose of diagnosis or treatment.
Heat unit means a unit of energy equal to the product of the peak kilovoltage, milliamperes, and seconds, i.e., kVp x mA x second.
HVL (See Half-value layer).
Image intensifier means a device, installed in its housing, which instantaneously converts an x-ray pattern into a corresponding light image of higher energy density.
Image receptor means any device, such as a fluorescent screen or radiographic film, which transforms incident x-ray photons either into a visible image or into another form which can be made into a visible image by further transformations.
Image receptor support means, for mammographic systems, that part of the system designed to support the image receptor in a horizontal plane during a mammographic examination.
Inherent filtration means the filtration of the useful beam provided by the permanently installed components of the tube housing assembly.
Irradiation means the exposure of matter to ionizing radiation.
Kilovolts peak (See Peak tube potential).
kV means kilovolts.
kVp (See Peak tube potential).
kWs means kilowatt second. It is equivalent to 103 kV.mA.s, i.e.,
(A)kWs = (X)kV x (Y)mA x (Z)s x kWs = XYZ kWs
103kV x mA x s 103
Lead equivalent means the thickness of lead affording the same attenuation, under specified conditions, as the material in question.
Leakage radiation means radiation emanating from the diagnostic or therapeutic source assembly except for:
Leakage technique factors means the technique factors associated with the diagnostic or therapeutic source assembly which are used in measuring leakage radiation. They are defined as follows:
Light field means that area of the intersection of the light beam from the beam-limiting device and one of the set of planes parallel to and including the plane of the image receptor, whose perimeter is the locus of points at which the illumination is one-fourth of the maximum in the intersection.
Linear attenuation coefficient or µ means the quotient of dN/N divided by dl when dN/N is the fraction of uncharged ionizing radiation that experience interactions in traversing a distance dl in a specified material.
Line-voltage regulation means the difference between the no-load and the load line potentials expressed as a percent of the load line potential. It is calculated using the following equation:
Percent line-voltage regulation = 100 (Vn-Vl)/Vl
where
Vn = No-load line potential and
Vl = Load line potential.
mAs means milliampere second.
Maximum line current means the root-mean-square current in the supply line of an x-ray machine operating at its maximum rating.
Mobile x-ray equipment (See X-ray equipment).
Patient means an individual subjected to healing arts examination, diagnosis, or treatment.
Peak tube potential means the maximum value of the potential difference across the x-ray tube during an exposure.
Phantom means a volume of material behaving in a manner similar to tissue with respect to the attenuation and scattering of radiation.
Phototimer means a method for controlling radiation exposures to image receptors by the amount of radiation which reaches a radiation monitoring device(s). The radiation monitoring device(s) is part of an electronic circuit which controls the duration of time the tube is activated (See Automatic exposure control).
PID (See Position indicating device).
Portable x-ray equipment (See X-ray equipment).
Position indicating device means a device on dental x-ray equipment used to indicate the beam position and to establish a definite source-surface (skin) distance. It may or may not incorporate or serve as a beam-limiting device.
Primary dose monitoring system means a system which will monitor the useful beam during irradiation and which will terminate irradiation when a pre-selected number of dose monitor units have been acquired.
Primary protective barrier (See Protective barrier).
Protective apron means an apron made of radiation absorbing materials used to reduce radiation exposure.
Protective barrier means a barrier of radiation absorbing material(s) used to reduce radiation exposure. The types of protective barriers are as follows:
Protective glove means a glove made of radiation absorbing materials used to reduce radiation exposure.
Qualified expert means an individual who has demonstrated to the satisfaction of the Agency that such individual possesses the knowledge and training to measure ionizing radiation, to evaluate safety techniques, and to advise regarding radiation protection needs.
Radiation detector means a device which in the presence of radiation provides a signal or other indication suitable for use in measuring one or more quantities of incident radiation.
Radiation therapy simulation system means a radiographic or fluoroscopic x-ray system intended for localizing the volume to be exposed during radiation therapy and confirming the position and size of the therapeutic irradiation field.
Radiograph means an image receptor on which the image is created directly or indirectly by an x-ray pattern and results in a permanent record.
Radiographic imaging system means any system whereby a permanent or semi-permanent image is recorded on an image receptor by the action of ionizing radiation.
Radiological physicist means an individual who,
Rating means the operating limits as specified by the component manufacturer.
Recording means producing a permanent form of an image resulting from x-ray photons.
Response time means the time required for an instrument system to reach 90 percent of its final reading when the radiation-sensitive volume of the instrument system is exposed to a step change in radiation flux from zero sufficient to provide a steady state midscale reading.
Scattered radiation means radiation that, during passage through matter, has been deviated in direction (See Direct scattered radiation).
Secondary dose monitoring system means a system which will terminate irradiation in the event of failure of the primary system.
Secondary protective barrier (See Protective barrier).
Shutter means a device attached to the tube housing assembly which can totally intercept the useful beam and which has a lead equivalency not less than that of the tube housing assembly.
SID (see Source-image receptor distance).
Source means the focal spot of the x-ray tube.
Source-image receptor distance means the distance from the source to the center of the input surface of the image receptor.
Spot check means a procedure which is performed to assure that a previous calibration continues to be valid.
Spot film means a radiograph which is made during a fluoroscopic examination to permanently record conditions which exist during that fluoroscopic procedure.
Spot-film device means a device intended to transport and/or position a radiographic image receptor between the x-ray source and fluoroscopic image receptor. It includes a device intended to hold a cassette over the input end of an image intensifier for the purpose of making a radiograph.
SSD means the distance between the source and the skin of the patient.
Stationary x-ray equipment (See X-ray equipment).
Stray radiation means the sum of leakage and scattered radiation.
Technique factors means the following conditions of operation:
Termination of irradiation means the stopping of irradiation in a fashion which will not permit continuance of irradiation without the resetting of operating conditions at the control panel.
Tomogram means the depiction of the x-ray attenuation properties of a section through the body.
Traceable to a national standard means that a quantity or a measurement has been compared to a national standard directly or indirectly through one or more intermediate steps and that all comparisons have been documented.
Tube means an x-ray tube, unless otherwise specified.
Tube housing assembly means the tube housing with tube installed. It includes high-voltage and/or filament transformers and other appropriate elements when such are contained within the tube housing.
Tube rating chart means the set of curves which specify the rated limits of operation of the tube in terms of the technique factors.
Useful beam means the radiation emanating from the tube housing port or the radiation head and passing through the aperture of the beam limiting device when the exposure controls are in a mode to cause the system to produce radiation.
Variable-aperture beam-limiting device means a beam-limiting device which has capacity for stepless adjustment of the x-ray field size at a given SID.
Visible area means that portion of the input surface of the image receptor over which incident x-ray photons are producing a visible image.
Wedge filter means an added filter effecting continuous progressive attenuation on all or part of the useful beam.
X-ray control means a device which controls input power to the x-ray high-voltage generator and/or the x-ray tube. It includes equipment such as timers, phototimers, automatic brightness stabilizers, and similar devices, which control the technique factors of an x-ray exposure.
X-ray equipment means an x-ray system, subsystem, or component thereof. Types of x-ray equipment are as follows:
X-ray field means that area of the intersection of the useful beam and any one of the set of planes parallel to and including the plane of the image receptor, whose perimeter is the locus of points at which the exposure rate is one-fourth of the maximum in the intersection.
X-ray high-voltage generator means a device which transforms electrical energy from the potential supplied by the x-ray control to the tube operating potential. The device may also include means for transforming alternating current to direct current, filament transformers for the x-ray tube(s), high-voltage switches, electrical protective devices, and other appropriate elements.
X-ray subsystem means any combination of two or more components of an x-ray system.
X-ray system means an assemblage of components for the controlled production of x rays. It includes minimally an x-ray high-voltage generator, an x-ray control, a tube housing assembly, a beam-limiting device, and the necessary supporting structures. Additional components which function with the system are considered integral parts of the system.
X-ray tube means any electron tube which is designed to be used primarily for the production of x rays.
1/This test is described by Suleiman, O.H. et al. in the article "Automatic Film Processing: Analysis of 9 Years of Observations." Radiology 1992 Vol. 185, pp. 25-28.
manufacturer). Solution may be removed from the tank to permit the addition of an adequate volume of replenisher.
No person shall modify a radiation machine, or any other auxiliary equipment that functions with the radiation machine to produce the result desired by use of the machine, in such a manner that the machine or auxiliary equipment fails to operate properly or otherwise does not meet any provision of these regulations.
| Design Operating Range (kVp) | Measured Potential (kVp) | Half-value layer (mm of aluminum) |
| Below 50 | 30 | 0.3 |
| Below 50 | 40 | 0.4 |
| Below 50 | 49 | 0.5 |
| 50 to 70 | 50 | 1.2 |
| 50 to 70 | 60 | 1.3 |
| 50 to 70 | 70 | 1.5 |
| Above 70 | 71 | 2.1 |
| Above 70 | 80 | 2.3 |
| Above 70 | 90 | 2.5 |
| Above 70 | 100 | 2.7 |
| Above 70 | 110 | 3.0 |
| Above 70 | 120 | 3.2 |
| Above 70 | 130 | 3.5 |
| Above 70 | 140 | 3.8 |
| Above 70 | 150 | 4.1 |
T % 5 (Tmax - Tmin).
_ _ _ _
%Y1 - Y2% % 0.20 (Y1 + Y2),
_ _
where Y1 - Y2 are the average mR/mAs (microcoulomb/kilogram per mAs) values obtained at each of 2 consecutive tube current settings.
_
E % 5(Emax - Emin).
%X1 - X2% % 0.10 (X1 + X2),
_ _
where X 1 and X 2 are the average mR/mAs (microcoulomb/kilogram per mAs) values obtained at each of 2 consecutive tube current settings.
_
T % 5(Tmax - Tmin).
E % 5(Emax - Emin).
%X1 - X2% % 0.10 (X1 + X2),
_ _
where X 1 and X 2 are the average mR/mAs values obtained at each of 2 consecutive tube current settings.
In addition to the definitions provided in A.2 and F.2 of these regulations, the following definitions shall be applicable to F.11:
Computed tomography dose index means the integral from -7T to +7T of the dose profile along a line perpendicular to the tomographic plane divided by the product of the nominal tomographic section thickness and the number of tomograms produced in a single scan, that is:
1 %+7T
CTDI = %-7T D(z)dz
nT
where:
z = Position along a line perpendicular to the tomographic plane.
D(z) = Dose at position z.
T = Nominal tomographic section thickness.
n = Number of tomograms produced in a single scan.
This definition assumes that the dose profile is centered around z=0 and that, for a multiple tomogram system, the scan increment between adjacent scans is nT.
Contrast scale means the change in the linear attenuation coefficient per CTN relative to water, that is:
µx - µw CS =
(CTN)x - (CTN)w
where:
µx = Linear attenuation coefficient of the material of interest.
µw = Linear attenuation coefficient of water.
(CTN)x = CTN of the material of interest.
(CTN)w = CTN of water.
CS (See Contrast scale).
CT conditions of operation" means all selectable parameters governing the operation of a CT x-ray system including, but not limited to, nominal tomographic section thickness, filtration, and the technique factors as defined in F.2.
CTDI (See Computed tomography dose index).
CT gantry means the tube housing assemblies, beam-limiting devices, detectors, and the supporting structures and frames which hold these components.
CTN (See CT number).
CT number means the number used to represent the x-ray attenuation associated with each elemental area of the CT image.
k( µx - µw)
CTN =
µw
where:
k = A constant [5/]
µx = Linear attenuation coefficient of the material of interest.
µw = Linear attenuation coefficient of water.
Dose profile means the dose as a function of position along a line.
Elemental area means the smallest area within a tomogram for which the x-ray attenuation properties of a body are depicted. (See also Picture element).
Multiple tomogram system means a computed tomography x-ray system which obtains x-ray transmission data simultaneously during a single scan to produce more than one tomogram.
Noise means the standard deviation of the fluctuations in CTN expressed as a percentage of the attenuation coefficient of water. Its estimate (Sn) is calculated using the following expression:
Sn = 100 x CS x s
µw
5/ The constant has a normal value of 1,000 when the Houndsfield scale of CTN is used.
where: CS = Contrast scale
µw = Linear attenuation coefficient of water.
s = Estimated standard deviation of the CTN of picture elements in a specified area of the CT image.
Nominal tomographic section thickness means the full width at half-maximum of the sensitivity profile taken at the center of the cross-sectional volume over which x-ray transmission data are collected.
Picture element means an elemental area of a tomogram.
Reference plane means a plane which is displaced from and parallel to the tomographic plane.
Scan means the complete process of collecting x-ray transmission data for the production of a tomogram. Data can be collected simultaneously during a single scan for the production of one or more tomograms.
Scan increment means the amount of relative displacement of the patient with respect to the CT x-ray system between successive scans measured along the direction of such displacement.
Scan sequence means a preselected set of two or more scans performed consecutively under preselected CT conditions of operation.
Scan time means the period of time between the beginning and end of x-ray transmission data accumulation for a single scan.
Single tomogram system means a CT x-ray system which obtains x-ray transmission data during a scan to produce a single tomogram.
Tomographic plane means that geometric plane which is identified as corresponding to the output tomogram.
Tomographic section means the volume of an object whose x-ray attenuation properties are imaged in a tomogram.
6/ For the purpose of determining the CTDI, the manufacturer's statement as to the nominal tomographic section thickness for that particular system may be utilized.
Analytical x-ray equipment means equipment used for x-ray diffraction or fluorescence analysis.
Analytical x-ray system means a group of components utilizing x or gamma rays to determine the elemental composition or to examine the microstructure of materials.
Fail-safe characteristics mean a design feature which causes beam port shutters to close, or otherwise prevents emergence of the primary beam, upon the failure of a safety or warning device.
Local components mean part of an analytical x-ray system and include areas that are struck by x rays such as radiation source housings, port and shutter assemblies, collimators, sample holders, cameras, goniometers, detectors, and shielding, but do not include power supplies, transformers, amplifiers, readout devices, and control panels.
Normal operating procedures mean step-by-step instructions necessary to accomplish the analysis. These procedures shall include sample insertion and manipulation, equipment alignment, routine maintenance by the registrant [or licensee], and data recording procedures, which are related to radiation safety.
Open-beam configuration means an analytical x-ray system in which an individual could accidentally place some part of his body in the primary beam path during normal operation.
Primary beam means radiation which passes through an aperture of the source housing by a direct path from the x-ray tube or a radioactive source located in the radiation source housing.
A device which prevents the entry of any portion of an individual's body into the primary x-ray beam path or which causes the beam to be shut off upon entry into its path shall be provided on all open-beam configurations. A registrant or licensee may apply to the Agency for an exemption from the requirement of a safety device. Such application shall include:
Unused ports on radiation source housings shall be secured in the closed position in a manner which will prevent casual opening.
All analytical x-ray equipment shall be labeled with a readily discernible sign or signs bearing the radiation symbol and the words:
On open-beam configurations installed after October 10, 1982 each port on the radiation source housing shall be equipped with a shutter that cannot be opened unless a collimator or a coupling has been connected to the port.
Each radiation source housing shall be subject to the following requirements:
Each x-ray generator shall be supplied with a protective cabinet which limits leakage radiation measured at a distance of 5 centimeters from its surface such that it is not capable of producing a dose in excess of 0.25 millirem (2.5 µSv) in one hour.
The local components of an analytical x-ray system shall be located and arranged and shall include sufficient shielding or access control such that no radiation levels exist in any area surrounding the local component group which could result in a dose to an individual present therein in excess of the dose limits given in Part D of these regulations. For systems utilizing x-ray tubes, these levels shall be met at any specified tube rating
Each area or room containing analytical x-ray equipment shall be conspicuously posted with a sign or signs bearing the radiation symbol and the words "Caution - X-Ray Equipment" or words having a similar intent in accordance with Part D of these regulations.
Normal operating procedures shall be written and available to all analytical x-ray equipment workers. No individual shall be permitted to operate analytical x-ray equipment in any manner other than that specified in the procedures unless such individual has obtained written approval of the radiation safety officer.
No individual shall bypass a safety device or interlock unless such individual has obtained the approval of the radiation safety officer. Such approval shall be for a specified period of time. When a safety device or interlock has been bypassed, a readily discernible sign bearing the words "Safety Device Not Working", or words having a similar intent, shall be placed on the radiation source housing.
Except as specified in H.5(b), no operation involving removal of covers, shielding materials or tube housings or modifications to shutters, collimators, or beam stops shall be performed without ascertaining that the tube is off and will remain off until safe conditions have been restored. The main switch, rather than interlocks, shall be used for routine shutdown in preparation for repairs.
Radioactive source housings shall be opened for source replacement, leak testing, or other maintenance or repair procedures only by individuals authorized to specifically conduct such procedures under a license issued by the U.S. Nuclear Regulatory Commission, an Agreement State, or a Licensing State.
No individual shall be permitted to operate or maintain analytical x-ray equipment unless such individual has received instruction in and demonstrated competence as to:
"This report is furnished to you under the provisions of COMAR 26.12.01.01 Part J. You should preserve this report for further reference."
Sec. J.15 Consultation with Workers During Inspections
|
We would greatly appreciate your feedback about this site. Copyright © 2009 University of Maryland DES |
|
UM Home | Directories | Search | Calendar Maintained by Department of Environmental Safety Direct questions and comments to safety@umd.edu |